OPERATING AND SERVICE MANUAL MDS E-SCAN Monitor Divert E-Scan System Designed and Manufactured by THE ELECTRON MACHINE CORPORAT
MDS E-SCAN MPR E-SCAN /INSTALLATION 10The adapter should be oriented to place the head in a horizontal plane to as
MDS E-SCAN MPR E-SCAN /INSTALLATION 11It is essential that NO LEAKS be present in this line that could allow evapo
MDS E-SCAN MPR E-SCAN /INSTALLATION 121.5 Cooling System Requirements When supplied with a vortex cooler a clea
MDS E-SCAN MPR E-SCAN OPERATION 13 2. MPR E-Scan Operation 2.1 The Operator’s Panel The Operator’s Panel or Front Panel consists of a di
MDS E-SCAN MPR E-SCAN OPERATION 142.1.2 Button Touch Pad Function Chart Scrolls cursor horizontally to the left and also used to initiate e
MDS E-SCAN MPR E-SCAN OPERATION 152.2 Normal Mode 2.2.1 What Normal Mode Does In normal mode the system displays the following informati
MDS E-SCAN MPR E-SCAN OPERATION 16Note: The screen names and set points below are just an example, they will vary from application to applic
MDS E-SCAN MPR E-SCAN OPERATION 17 2.2.4 Overview of Menu Options CALIB (Password required) option allows the operator to enter calibra
MDS E-SCAN MPR E-SCAN OPERATION/CALIBRATION 18 2.3 Calibration Selection (Note: To access calibration enter password “0000") The calib
MDS E-SCAN MPR E-SCAN OPERATION/CALIBRATION 19 Measurement Alarm Divert: If the measurement falls below this limit, a system alarm is activa
MDS E-SCAN MPR E-SCAN OPERATION/CALIBRATION 20 Min. Purge Temperature: This value determines the point below which a steam purge will not
MDS E-SCAN MPR E-SCAN OPERATION/CALIBRATION 21 Selecting the “Temp.” option from the calibration menu results in the following display: (Cir
MDS E-SCAN MPR E-SCAN OPERATION/CALIBRATION 22 The final option of the calibration menu is “Anlg. 0" which activates the following display:
MDS E-SCAN MPR E-SCAN OPERATION/CALIBRATION 23 Note: If the process measurement changes after the sampled reading is recorded then the final adj
MDS E-SCAN MPR E-SCAN OPERATION/ DIAGNOSITIC SELECTION 24 2.5 Diagnostic Selection The Diagnostics menu contains the following options: 1.
MDS E-SCAN MPR E-SCAN OPERATION/ DIAGNOSITIC SELECTION 25 The “Relays” option activates the following display: There are eight total relays,
MDS E-SCAN MPR E-SCAN OPERATION/ DIAGNOSITIC SELECTION 26 NOTE: The information displayed at the bottom gives the operator the output resolution
MDS E-SCAN MPR E-SCAN OPERATION/ PURGE SELECTION 27 During the purge process, all readings and analog output are “on hold” or frozen. 2.6 Purge
MDS E-SCAN MPR E-SCAN OPERATION/ PURGE SELECTION 28 2.7 Configuration Section Display Unit Number This is used to identify refract #1 or #
MDS E-SCAN MPR E-SCAN OPERATION / PROBLEM ANAYLSIS 29 3. MPR E-Scan Problem Analysis 3.1 Procedures The following analysis procedures ar
The Electron Machine Corporation MDS E-SCAN Introduction Increasing emphasis on recovery boiler safety in pulp and paper manufacture requir
MDS E-SCAN MPR E-SCAN OPERATION / PROBLEM ANAYLSIS 30 3.2 Trouble Shooting Chart SYMPTOM POSSIBLE CAUSE No display, failsafe LED off. No d
MDS E-SCAN MPR E-SCAN INFORMATION 31 4. MPR E-Scan Information 4.1 Technical Description 4.1.1 Sensing Head The MPR E-Scan uses an
MDS E-SCAN MPR E-SCAN INFORMATION 32 4.1.1.2 CCD Linear Array The linear array used in the E-Scan includes 2048 individual photo sites.
MDS E-SCAN MPR E-SCAN INFORMATION 33 The signal is passed through circuitry to develop a DC voltage, which is related to the ratio of the time
MDS E-SCAN MPR E-SCAN INFORMATION 34 4.1.2.3 IB (interface board) for external connection The interface board is located on the back wall
MDS E-SCAN MPR E-SCAN INFORMATION 35 4.1.2.7 CPU Card The CPU card is found plugged into the PC/104 bus connectors of the I/O card. The CP
MDS E-SCAN MPR E-SCAN INFORMATION 36 4.2 Refractometer Specifications Accuracy +/- 1% of span or .000075 R.I., whichever is gre
MDS E-SCAN MONITOR OPERATION 37 5. Monitor Operation 5.1 The Operator’s Panel The Operator’s Panel or Front Panel consists of
MDS E-SCAN MONITOR OPERATION 38 5.1.2 Button Touch Pad Function Chart Scrolls cursor horizontally to the left.Scrolls cursor horizo
MDS E-SCAN MONITOR OPERATION 39 Note: The readings below are just an example, they will vary from application to application. 5.2 N
CAUTION When removing the sensing head from an operating line, do not assume that the line is empty or that the isolation or bypass means is w
MDS E-SCAN MONITOR OPERATION 40 +15 VOLT ERROR: This indicates that the plus 15 volt supply is outside the allowable 10% tolerance l
MDS E-SCAN MONITOR OPERATION 41 PURGE MISSED: This indicates that the steam solenoid relay was not activated during the purge operat
MDS E-SCAN MONITOR OPERATION 42 SYSTEM: This status field is used to monitor many system functions and to issue warnings instead of
MDS E-SCAN MONITOR OPERATION 43 Note: The readings below are just an example, they will vary from application to application. 5.2.4
MDS E-SCAN MONITOR LOG SELECTION 44 5.3 Log Selection The logs menu contains four menu options: 1. ERR/WARN LOG This option allows th
MDS E-SCAN MONITOR CONFIGURATION SELECTION 45 5.4 Configuration Selection The configuration menu contains two menu options: 1. STATES
MDS E-SCAN MONITOR CONFIGURATION STATES SELECTION 46 5.5 Configuration States Selection The first configuration option, “States”, activa
MDS E-SCAN MONITOR CONFIGURATION LIMITS SELECTION 47 5.6 Configuration Limits Selection The next configuration option, “Limits”, activa
MDS E-SCAN MONITOR DIAGNOSTICS MODE SELECTION 48 5.7 Diagnostic Mode Selection The Diagnostics menu contains the following options: 1
MDS E-SCAN MONITOR DIAGNOSTIC VOLTAGES SELECTION 49 5.8 Diagnostic Voltages Selection The “Voltages” option results in the following disp
Table of Contents 1. MDS E-Scan Installation... 9 1.1 Site Selection...
MDS E-SCAN MONITOR DIAGNOSTIC VOLTAGES SELECTION 50 U2 Meas: The voltage produced when measuring a solution sample. It is used by MPR E-Sca
MDS E-SCAN MONITOR DIAGNOSTIC RELAY SELECTION 51 5.9 Diagnostic Relay Selection The “Relays” option activates the following display: The
MDS E-SCAN MONITOR DIAGNOSTIC TIME/DATE SELECTION 52 5.10 Diagnostic Time/Date Selection The “Time/Date” diagnostic option activates the fo
MDS E-SCAN MONITOR PROBLEM ANALYSIS 53 6. Monitor Problem Analysis 6.1 Procedures The following analysis procedures are meant to aid i
MDS E-SCAN MONITOR INFORMATION 54 7. Monitor Information 7.1 Console Technical Description The console is constructed of molded fibe
MDS E-SCAN MONITOR INFORMATION 55 7.1.1 Console Cards IB (interface board) for external connection The interface board is located on the b
MDS E-SCAN MONITOR INFORMATION 56 7.3 Features • Continuously monitors both refractometers and associated equipment, looking at all vital
MDS E-SCAN MONITOR INFORMATION 57 7.4 Specific Functions (Dwg. A-11585) Note: 58% +2% noted below are industry standards but may vary de
MDS E-SCAN MONITOR INFORMATION 58 7. Monitor the isolation valve position. A partially closed or closed valve will initiate steps (a), (b)
MDS E-SCAN MONITOR INFORMATION 59 7.5 Trouble Shooting Guide and Truth Table (For unit configured for two refractometer divert.) (X) Refr
4. MPR E-Scan Information...31 4.1 Technical Description...
MDS E-SCAN MONITOR INFORMATION 60 Condition for either Refractometer (1) or (2) Alarms Activated Liquor Diversion Additional Notes Loss
MDS E-SCAN MONITOR INFORMATION 61 7.5.2 Warning Alarm Conditions Alarm Condition 8 System Warning. * Check lamp voltage, indicates la
MDS E-SCAN SERVICE 62 Please include model number and serial number in all correspondence. 8. Service 8.1 EMC Warranty The Electron Mac
MDS E-SCAN SERVICE 63 Purchase orders should include the model number and the serial number of the equipment that the part is to be used with, i
MDS E-SCAN SERVICE 64 8.5 Maintenance MDS E-Scan Monitor Divert System Periodic Inspection Checklist We recommend the following checks be pe
MDS E-SCAN SERVICE 65 8.6 Caution CAUTION When removing the sensing head from an operating line, do not assume that the line is empty or t
MDS E-SCAN SERVICE 66 -Notes-
MDS E-SCAN DRAWING LIST 67 9. Drawings List Drawing # Rev Description A-11575 B Outline, MDS E-Scan Hoffman Console B-11586 H MDS E-
6. Monitor Problem Analysis...53 6.1 Procedures...
MDS E-SCAN MPR E-SCAN /INSTALLATION 9MDS E-Scan NEMA 4X Console 1. MDS E-Scan Installation 1.1 Site Selec