Página 1 - MDS E-SCAN

OPERATING AND SERVICE MANUAL MDS E-SCAN Monitor Divert E-Scan System Designed and Manufactured by THE ELECTRON MACHINE CORPORAT

Página 2

MDS E-SCAN MPR E-SCAN /INSTALLATION 10The adapter should be oriented to place the head in a horizontal plane to as

Página 3 - Introduction

MDS E-SCAN MPR E-SCAN /INSTALLATION 11It is essential that NO LEAKS be present in this line that could allow evapo

Página 4 - Important Manual Information

MDS E-SCAN MPR E-SCAN /INSTALLATION 121.5 Cooling System Requirements When supplied with a vortex cooler a clea

Página 5

MDS E-SCAN MPR E-SCAN OPERATION 13 2. MPR E-Scan Operation 2.1 The Operator’s Panel The Operator’s Panel or Front Panel consists of a di

Página 6

MDS E-SCAN MPR E-SCAN OPERATION 142.1.2 Button Touch Pad Function Chart Scrolls cursor horizontally to the left and also used to initiate e

Página 7

MDS E-SCAN MPR E-SCAN OPERATION 152.2 Normal Mode 2.2.1 What Normal Mode Does In normal mode the system displays the following informati

Página 8

MDS E-SCAN MPR E-SCAN OPERATION 16Note: The screen names and set points below are just an example, they will vary from application to applic

Página 9 - 1.2 Power Requirements

MDS E-SCAN MPR E-SCAN OPERATION 17 2.2.4 Overview of Menu Options CALIB (Password required) option allows the operator to enter calibra

Página 10

MDS E-SCAN MPR E-SCAN OPERATION/CALIBRATION 18 2.3 Calibration Selection (Note: To access calibration enter password “0000") The calib

Página 11 - 1.4 Steam Purge Attachment

MDS E-SCAN MPR E-SCAN OPERATION/CALIBRATION 19 Measurement Alarm Divert: If the measurement falls below this limit, a system alarm is activa

Página 13 - 2. MPR E-Scan Operation

MDS E-SCAN MPR E-SCAN OPERATION/CALIBRATION 20 Min. Purge Temperature: This value determines the point below which a steam purge will not

Página 14

MDS E-SCAN MPR E-SCAN OPERATION/CALIBRATION 21 Selecting the “Temp.” option from the calibration menu results in the following display: (Cir

Página 15 - 2.2 Normal Mode

MDS E-SCAN MPR E-SCAN OPERATION/CALIBRATION 22 The final option of the calibration menu is “Anlg. 0" which activates the following display:

Página 16

MDS E-SCAN MPR E-SCAN OPERATION/CALIBRATION 23 Note: If the process measurement changes after the sampled reading is recorded then the final adj

Página 17

MDS E-SCAN MPR E-SCAN OPERATION/ DIAGNOSITIC SELECTION 24 2.5 Diagnostic Selection The Diagnostics menu contains the following options: 1.

Página 18

MDS E-SCAN MPR E-SCAN OPERATION/ DIAGNOSITIC SELECTION 25 The “Relays” option activates the following display: There are eight total relays,

Página 19

MDS E-SCAN MPR E-SCAN OPERATION/ DIAGNOSITIC SELECTION 26 NOTE: The information displayed at the bottom gives the operator the output resolution

Página 20

MDS E-SCAN MPR E-SCAN OPERATION/ PURGE SELECTION 27 During the purge process, all readings and analog output are “on hold” or frozen. 2.6 Purge

Página 21

MDS E-SCAN MPR E-SCAN OPERATION/ PURGE SELECTION 28 2.7 Configuration Section Display Unit Number This is used to identify refract #1 or #

Página 22

MDS E-SCAN MPR E-SCAN OPERATION / PROBLEM ANAYLSIS 29 3. MPR E-Scan Problem Analysis 3.1 Procedures The following analysis procedures ar

Página 23 - 2.4 Print Selection

The Electron Machine Corporation MDS E-SCAN Introduction Increasing emphasis on recovery boiler safety in pulp and paper manufacture requir

Página 24 - 2.5 Diagnostic Selection

MDS E-SCAN MPR E-SCAN OPERATION / PROBLEM ANAYLSIS 30 3.2 Trouble Shooting Chart SYMPTOM POSSIBLE CAUSE No display, failsafe LED off. No d

Página 25

MDS E-SCAN MPR E-SCAN INFORMATION 31 4. MPR E-Scan Information 4.1 Technical Description 4.1.1 Sensing Head The MPR E-Scan uses an

Página 26

MDS E-SCAN MPR E-SCAN INFORMATION 32 4.1.1.2 CCD Linear Array The linear array used in the E-Scan includes 2048 individual photo sites.

Página 27 - 2.6 Purge Selection

MDS E-SCAN MPR E-SCAN INFORMATION 33 The signal is passed through circuitry to develop a DC voltage, which is related to the ratio of the time

Página 28

MDS E-SCAN MPR E-SCAN INFORMATION 34 4.1.2.3 IB (interface board) for external connection The interface board is located on the back wall

Página 29 - 3.1 Procedures

MDS E-SCAN MPR E-SCAN INFORMATION 35 4.1.2.7 CPU Card The CPU card is found plugged into the PC/104 bus connectors of the I/O card. The CP

Página 30 - 3.2 Trouble Shooting Chart

MDS E-SCAN MPR E-SCAN INFORMATION 36 4.2 Refractometer Specifications Accuracy +/- 1% of span or .000075 R.I., whichever is gre

Página 31 - 4. MPR E-Scan Information

MDS E-SCAN MONITOR OPERATION 37 5. Monitor Operation 5.1 The Operator’s Panel The Operator’s Panel or Front Panel consists of

Página 32

MDS E-SCAN MONITOR OPERATION 38 5.1.2 Button Touch Pad Function Chart Scrolls cursor horizontally to the left.Scrolls cursor horizo

Página 33 - Filtered CCD Signal

MDS E-SCAN MONITOR OPERATION 39 Note: The readings below are just an example, they will vary from application to application. 5.2 N

Página 34

CAUTION When removing the sensing head from an operating line, do not assume that the line is empty or that the isolation or bypass means is w

Página 35

MDS E-SCAN MONITOR OPERATION 40 +15 VOLT ERROR: This indicates that the plus 15 volt supply is outside the allowable 10% tolerance l

Página 36

MDS E-SCAN MONITOR OPERATION 41 PURGE MISSED: This indicates that the steam solenoid relay was not activated during the purge operat

Página 37 - 5. Monitor Operation

MDS E-SCAN MONITOR OPERATION 42 SYSTEM: This status field is used to monitor many system functions and to issue warnings instead of

Página 38

MDS E-SCAN MONITOR OPERATION 43 Note: The readings below are just an example, they will vary from application to application. 5.2.4

Página 39 - 5.2 Normal Mode

MDS E-SCAN MONITOR LOG SELECTION 44 5.3 Log Selection The logs menu contains four menu options: 1. ERR/WARN LOG This option allows th

Página 40

MDS E-SCAN MONITOR CONFIGURATION SELECTION 45 5.4 Configuration Selection The configuration menu contains two menu options: 1. STATES

Página 41

MDS E-SCAN MONITOR CONFIGURATION STATES SELECTION 46 5.5 Configuration States Selection The first configuration option, “States”, activa

Página 42

MDS E-SCAN MONITOR CONFIGURATION LIMITS SELECTION 47 5.6 Configuration Limits Selection The next configuration option, “Limits”, activa

Página 43

MDS E-SCAN MONITOR DIAGNOSTICS MODE SELECTION 48 5.7 Diagnostic Mode Selection The Diagnostics menu contains the following options: 1

Página 44 - 5.3 Log Selection

MDS E-SCAN MONITOR DIAGNOSTIC VOLTAGES SELECTION 49 5.8 Diagnostic Voltages Selection The “Voltages” option results in the following disp

Página 45 - 5.4 Configuration Selection

Table of Contents 1. MDS E-Scan Installation... 9 1.1 Site Selection...

Página 46

MDS E-SCAN MONITOR DIAGNOSTIC VOLTAGES SELECTION 50 U2 Meas: The voltage produced when measuring a solution sample. It is used by MPR E-Sca

Página 47

MDS E-SCAN MONITOR DIAGNOSTIC RELAY SELECTION 51 5.9 Diagnostic Relay Selection The “Relays” option activates the following display: The

Página 48

MDS E-SCAN MONITOR DIAGNOSTIC TIME/DATE SELECTION 52 5.10 Diagnostic Time/Date Selection The “Time/Date” diagnostic option activates the fo

Página 49

MDS E-SCAN MONITOR PROBLEM ANALYSIS 53 6. Monitor Problem Analysis 6.1 Procedures The following analysis procedures are meant to aid i

Página 50

MDS E-SCAN MONITOR INFORMATION 54 7. Monitor Information 7.1 Console Technical Description The console is constructed of molded fibe

Página 51

MDS E-SCAN MONITOR INFORMATION 55 7.1.1 Console Cards IB (interface board) for external connection The interface board is located on the b

Página 52 - 5.11 Print Selection

MDS E-SCAN MONITOR INFORMATION 56 7.3 Features • Continuously monitors both refractometers and associated equipment, looking at all vital

Página 53 - 6.2 Trouble Shooting Chart

MDS E-SCAN MONITOR INFORMATION 57 7.4 Specific Functions (Dwg. A-11585) Note: 58% +2% noted below are industry standards but may vary de

Página 54 - 7. Monitor Information

MDS E-SCAN MONITOR INFORMATION 58 7. Monitor the isolation valve position. A partially closed or closed valve will initiate steps (a), (b)

Página 55 - 7.2 Monitor Specifications

MDS E-SCAN MONITOR INFORMATION 59 7.5 Trouble Shooting Guide and Truth Table (For unit configured for two refractometer divert.) (X) Refr

Página 56 - 7.3 Features

4. MPR E-Scan Information...31 4.1 Technical Description...

Página 57 - 7.4 Specific Functions

MDS E-SCAN MONITOR INFORMATION 60 Condition for either Refractometer (1) or (2) Alarms Activated Liquor Diversion Additional Notes Loss

Página 58

MDS E-SCAN MONITOR INFORMATION 61 7.5.2 Warning Alarm Conditions Alarm Condition 8 System Warning. * Check lamp voltage, indicates la

Página 59

MDS E-SCAN SERVICE 62 Please include model number and serial number in all correspondence. 8. Service 8.1 EMC Warranty The Electron Mac

Página 60

MDS E-SCAN SERVICE 63 Purchase orders should include the model number and the serial number of the equipment that the part is to be used with, i

Página 61

MDS E-SCAN SERVICE 64 8.5 Maintenance MDS E-Scan Monitor Divert System Periodic Inspection Checklist We recommend the following checks be pe

Página 62 - 8. Service

MDS E-SCAN SERVICE 65 8.6 Caution CAUTION When removing the sensing head from an operating line, do not assume that the line is empty or t

Página 63 - 8.4 Spare Parts

MDS E-SCAN SERVICE 66 -Notes-

Página 64 - 8.5 Maintenance

MDS E-SCAN DRAWING LIST 67 9. Drawings List Drawing # Rev Description A-11575 B Outline, MDS E-Scan Hoffman Console B-11586 H MDS E-

Página 65 - 8.6 Caution

6. Monitor Problem Analysis...53 6.1 Procedures...

Página 67 - 9. Drawings List

MDS E-SCAN MPR E-SCAN /INSTALLATION 9MDS E-Scan NEMA 4X Console 1. MDS E-Scan Installation 1.1 Site Selec